Customers with questions or concerns regarding the recall notice, are instructed to contact Beckman Coulter Customer Support at or call (800) 526-7694 in US and Canada. Customers are instructed to complete and return the enclosed Response Form within 10 days. The letter informs the customers of the problems identified and the actions to be taken. As a result of internal investigations following customer complaints, Iris International has determined that the iChemVELOCITY systems do not maintain the on-board strip stability claim of 5 days at 18-28❼ at 20-80% relative humidity (RH) as stated in the iChemVELOCITY product labeling. Iris International is recalling the iChemVELOCITY systems because they do not maintain the on-board strip stability claim of 5 days at 18-28 degrees celsius at 20-80% relative humidity (RH) as stated in the iChemVELOCITY product labeling.Īn Urgent Medical Device Recall letter dated 7/15/15 was sent to all customers who purchased the iChemVELOCITY Automated Urine Chemistry System using iChemVELOCITY Urine Chemistry Strips REF 800-7212 and 800-7212-001.

beckman coulter chatsworth

An in vitro diagnostic device used to automate the urine chemistry analysis profile using iChemVELOCITY Urine Chemistry Strips. IChemVELOCITY Automated Urine Chemistry Catalog No.

#Beckman coulter chatsworth code

Class 2 Device Recall iChemVELOCITY Automated Urine Chemistry SystemĪutomated urinalysis system - Product Code KQO